The main ADC sampling rate and the rate at which the analog signal outputs are updated is 1 MSa/s, giving excellent performance when used at short output filter time constant settings, such as in scanned probe measurements. But weve also increased the maximum rate at which data can be stored to the internal curve buffer to 1 s per point, allowing for the first time direct capture of instrument outputs when using these short time constants. The buffer length has also been increased to 100,000 sets of points, giving recording times of 100 ms at the fastest sampling rates. Whats more, in the fast capture mode the length does not need to be divided by the number of outputs being stored, making it possible, for example, to store the full 100,000 points of X, Y and auxiliary ADC1 values at the same time.