| Est. Shipping Width Inches (mm) | 31 (787) |
| Est. Shipping Height Inches (mm) | 51(1219) |
| Est. Shipping Depth Inches (mm) | 31 (787) |
| Access Port Location | Right Side |
| Air Cleanliness (ISO 14644) | ISO-Class-5 |
| Chamber Air Circulation Method | Air Pump |
| Chamber Air Filter | 99.97% Efficiency @ &ge, 0.3 Micron Capsule Filter Located in Sensor Bay |
| Chamber Heater(s) Location | Sides, Bottom, Top and Back of Chamber |
| Chamber Jacket Technology | Air Jacket via Two Axial Fans Drawing Air into Shell |
| CO2 Accuracy | ±, 0.1% |
| CO2 Control Logic | Fixed Algorithm/Manual, Environmental Adaptable |
| CO2 Display Resolution | 0.1% |
| CO2 Range | 0.1 to 20%, 5% Default Set Point |
| CO2 Recovery | Up to 5% ±, 0.2% / -0.5 in 5 minutes Average |
| CO2 Sensor Type | Infrared (NDIR) |
| Control System | NuTouch |
| Data Output Features | RS-485, USB, 4-20mA |
| Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable |
| Electrical Configuration | 115V50Hz, 230V5060Hz |
| Exterior Depth Inches (mm) | 27.188 (691) |
| Exterior Door Seal | Magnetic Outer Door Gasket |
| Exterior Height Inches (mm) | 39.688 (1008) |
| Exterior Material | Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish |
| Exterior Width Inches (mm) | 26.750 (680) |
| Footprint Depth inches (mm) | 20.75 (527) |
| Footprint Width inches (mm) | 23.75 (603) |
| In-line Gas Filters | 99.97% Efficiency @ &ge, 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor |
| Segmented Doors | Optional Set of 4 Inner Doors |
| Inner Doors | One, single or sectional |
| Insulation | High Density Insulation with High R Rating |
| Interior Depth Inches (mm) | 20.053 (509) |
| Interior Door Seal | Silicone Rubber Gasket |
| Interior Height Inches (mm) | 28.500 (724) |
| Interior Material | Highly Polished Type 304L 16 Gauge Stainless Steel |
| Interior Volume Cubic Feet (Liters) | 7.06 (200) |
| Interior Width Inches (mm) | 21.257 (540) |
| Net Weight lbs (kg) | 235 (106) |
| O2 Accuracy | |
| O2 Control Logic | |
| O2 Display Resolution | |
| O2 Measuring Range | |
| O2 Set Point Range | |
| O2 Recovery | |
| O2 Sensor Type | |
| Power Cord Length Foot (Meter) | 8 (2.5) |
| Remote Alarm Terminals | Standard |
| RH Accuracy | |
| RH Control Method | Water Pan, Convection |
| RH Control Logic | |
| RH Control Range | |
| RH Display Resolution | |
| RH Evaporator Box Capacity | |
| RH Recovery | |
| RH Sensor Type | |
| RH Water Pan Capacity | |
| Sample Port Location | Inner Glass Door |
| Shelf Capacity lbs (kg) | |
| Shelf Depth Inches (mm) | 18 3/4 (476) |
| Shelf Material | Stainless Steel or Stainless Steel Enriched Copper |
| Shelf Width Inches (mm) | 19 3/4 (502) |
| Shelves | 4 Shelves Supplied, 20 Shelves Maximum Capacity |
| Est. Shipping Weight lbs (kg) | 325 (147) |
| Decontamination Cycle | 95C Humidified Decontamination, 145C Dry Decontamination Cycle |
| Temperature Accuracy C | ±, 0.1% |
| Temperature Control | Direct Heat |
| Temperature Control Range | 5C above ambient (to a 30C max ambient) to 55C |
| Temperature Display Resolution | 0.1C |
| Temperature Recovery | 0.12C/min. on Average |
| Temperature Sensor Type | Precision Integrated Circuit |
| Temperature Set-Point Range | 5C to 55C (37.0C Default) |
| Temperature Uniformity C | ±, 0.35C @ 37C |
| Water Pan Depth | 1.5 (38) |
| Water Pan Length Inches (mm) | 12 (305) |
| Water Pan Width | 10 (254) |