Каталог

Est. Shipping Width Inches (mm) 31 (787)
Est. Shipping Height Inches (mm) 51(1219)
Est. Shipping Depth Inches (mm) 31 (787)
Access Port Location Right Side
Air Cleanliness (ISO 14644) ISO-Class-5
Chamber Air Circulation Method Air Pump
Chamber Air Filter 99.97% Efficiency @ &ge, 0.3 Micron Capsule Filter Located in Sensor Bay
Chamber Heater(s) Location Sides, Bottom, Top and Back of Chamber
Chamber Jacket Technology Air Jacket via Two Axial Fans Drawing Air into Shell
CO2 Accuracy &plusmn, 0.1%
CO2 Control Logic Fixed Algorithm/Manual, Environmental Adaptable
CO2 Display Resolution 0.1%
CO2 Range 0.1 to 20%, 5% Default Set Point
CO2 Recovery Up to 5% &plusmn, 0.2% / -0.5 in 5 minutes Average
CO2 Sensor Type Infrared (NDIR)
Control System NuTouch
Data Output Features RS-485, USB, 4-20mA
Door and Perimeter Heater Control Logic Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable
Electrical Configuration 115V50Hz, 230V5060Hz
Exterior Depth Inches (mm) 27.188 (691)
Exterior Door Seal Magnetic Outer Door Gasket
Exterior Height Inches (mm) 39.688 (1008)
Exterior Material Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish
Exterior Width Inches (mm) 26.750 (680)
Footprint Depth inches (mm) 20.75 (527)
Footprint Width inches (mm) 23.75 (603)
In-line Gas Filters 99.97% Efficiency @ &ge, 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor
Segmented Doors Optional Set of 4 Inner Doors
Inner Doors One, single or sectional
Insulation High Density Insulation with High R Rating
Interior Depth Inches (mm) 20.053 (509)
Interior Door Seal Silicone Rubber Gasket
Interior Height Inches (mm) 28.500 (724)
Interior Material Highly Polished Type 304L 16 Gauge Stainless Steel
Interior Volume Cubic Feet (Liters) 7.06 (200)
Interior Width Inches (mm) 21.257 (540)
Net Weight lbs (kg) 235 (106)
O2 Accuracy
O2 Control Logic
O2 Display Resolution
O2 Measuring Range
O2 Set Point Range
O2 Recovery
O2 Sensor Type
Power Cord Length Foot (Meter) 8 (2.5)
Remote Alarm Terminals Standard
RH Accuracy
RH Control Method Water Pan, Convection
RH Control Logic
RH Control Range
RH Display Resolution
RH Evaporator Box Capacity
RH Recovery
RH Sensor Type
RH Water Pan Capacity
Sample Port Location Inner Glass Door
Shelf Capacity lbs (kg)
Shelf Depth Inches (mm) 18 3/4 (476)
Shelf Material Stainless Steel or Stainless Steel Enriched Copper
Shelf Width Inches (mm) 19 3/4 (502)
Shelves 4 Shelves Supplied, 20 Shelves Maximum Capacity
Est. Shipping Weight lbs (kg) 325 (147)
Decontamination Cycle 95C Humidified Decontamination, 145C Dry Decontamination Cycle
Temperature Accuracy C &plusmn, 0.1%
Temperature Control Direct Heat
Temperature Control Range 5C above ambient (to a 30C max ambient) to 55C
Temperature Display Resolution 0.1C
Temperature Recovery 0.12C/min. on Average
Temperature Sensor Type Precision Integrated Circuit
Temperature Set-Point Range 5C to 55C (37.0C Default)
Temperature Uniformity C &plusmn, 0.35C @ 37C
Water Pan Depth 1.5 (38)
Water Pan Length Inches (mm) 12 (305)
Water Pan Width 10 (254)
+79043698209